Telaire® 7001 Handheld CO2 Monitor
Equipped with Telaire's patented Dual Beam Absorption Infrared™ technology, the Telaire® is an easy-to-use CO2 / temperaure monitor designed for use in residential and commercial applications. With the ability to display CO2 readings in less then 30 seconds, the monitor is an ideal tool for: identifying energy saving opportunities in over-ventilated spaces, determining if air quality complaints are due to insufficient ventilation, or locating the presence of combustion fumes generated from vehicles and appliances. The Telaire® 7001 is also equipped with an output for recording data via an accessory cable and data logger. Recorded data can be graphed and printed using our Windows® based VG Graphing Software.
Specifications
|
|
Method
|
Dual Beam Absorption Infrared |
|
CO2 Channel
|
|
Measurement Range
|
0-4,000 ppm voltage output
0-10,000 ppm display |
|
Sensitivity
|
±1 ppm |
|
Accuracy
|
±50 ppm or 5% whichever is greater |
|
Repeatability
|
±20 ppm |
|
Temperature Dependence
|
±0.1% of reading per ºC or ±2 pmm per ºC, whichever is greater, referenced at 25ºC |
|
Pressure Dependence
|
0.13% of reading per mm Hg (Corrected via user inputfor elevation) |
|
Annual Drift
|
±20 ppm typical |
|
Response Time 0-90%
|
< 60 seconds for 90% of step change |
|
Warm-Up Time
|
< 60 seconds @ 22ºC |
|
Operating Conditions
|
0 - 50ºC (32º - 122ºF)
0-95% RH, non-condensing |
|
Storage Temperature
|
-40 - 60ºC (-40 - 140ºC) |
|
Calibration Interval
|
12 months, offset adjustment using single gas at 0-1,000 ppm CO2. Full factory calibration available |
|
Temperature Channel
|
|
Temperature Range
|
Voltage output - 0 - 40ºC (32 - 104ºF)
Display - 0 - 50ºC (32 - 122ºF) |
|
Display Resolution
|
0.1ºC (0.1ºF) |
|
Display Options
|
ºC, ºF, or off. Set with panel button |
|
Accuracy
|
±1ºC (2ºF) |
|
Response Time
|
20-30 minutes (case must equilibrate with environment) |
|
Calibration Interval
|
12 months, offset adjustment using temperature standard at 10 - 30ºC (50 - 86ºF)
Full factory calibration available |
|
Output - Analog
|
|
CO2
|
0-4VDC linear, 1mV / ppm (4,000 ppm max) |
|
Temperature
|
0-4VDC linear, 0 - 40ºC (32 - 104ºF) |
|
Output Impedance
|
100 Ohms |
|
Wiring Connection
|
Via RJ-45 to DB9 serial data cable |
|
Response Time
|
20-30 minutes (case must equilibrate with environment) |
|
Display
|
LCD with independent CO2 / temperature readings (panel buttons set elevation, ºC / ºF, calibration functions) |
|
Power Supply
|
|
Battery Type
|
Four AA batteries, not included |
|
|
|
|
Battery Opertion
|
80 hours (alkaline) |
|
External
|
6VDC from external AC/DC adapter, included |
|
Power Requirements
|
100mA peak, 20mA average from 6VDC |
|
Certification
|
FCC Class 15 Part B |
Downloads
|
Datasheet
Manual
|
|
|
|